CVE-2018-4846

CVE-2018-4846

Vendor Siemens Ag
Product RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems
Weakness CWE-798 · Hardcoded credentials
Published June 26, 2018
Last update September 17, 2024

CVSS base score

What the vulnerability does

01Description

A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

Key dates

02Disclosure timeline

June 26, 2018 CVE published
September 17, 2024 Record updated