CVE-2023-0888 MEDIUM

CVE-2023-0888: Authenticated eval injection in B. Braun Space Battery pack SP with Wi-Fi

Vendor B. Braun
Product Space Battery Pack SP with Wi-Fi
Weakness CWE-95 · Eval injection
Published March 13, 2023
Last update February 27, 2025

CVSS base score

4.9/10
Attack vector Network
Attack complexity Low
Privileges required High
User interaction None
Confidentiality None
Integrity High

CVSS vector

CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:N/I:H/A:N

What the vulnerability does

01Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Key dates

02Disclosure timeline

March 13, 2023 CVE published
February 27, 2025 Record updated