CVE-2023-30561 MEDIUM

CVE-2023-30561: Lack of Cryptographic Security of IUI Bus

Vendor Becton Dickinson & Co
Product BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015
Weakness CWE-311 · Missing encryption
Published July 13, 2023
Last update October 22, 2024

CVSS base score

6.1/10
Attack vector Physical
Attack complexity Low
Privileges required None
User interaction None
Confidentiality High
Integrity None

CVSS vector

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:H

What the vulnerability does

01Description

The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.

Key dates

02Disclosure timeline

July 13, 2023 CVE published
October 22, 2024 Record updated