CVE-2024-48966 CRITICAL

CVE-2024-48966: Life2000 service tools for test and calibration do not support user authentication

Vendor Baxter
Product Life2000 Ventilation System
Weakness CWE-306 · Missing auth
Published November 14, 2024
Last update November 15, 2024

CVSS base score

10.0/10
Attack vector Network
Attack complexity Low
Privileges required None
User interaction None
Confidentiality High
Integrity High

CVSS vector

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

What the vulnerability does

01Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Key dates

02Disclosure timeline

November 14, 2024 CVE published
November 15, 2024 Record updated