CVE-2025-0683 HIGH

CVE-2025-0683: Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

Vendor Contec Health
Product CMS8000 Patient Monitor
Weakness CWE-359
Published January 30, 2025
Last update February 12, 2025

CVSS base score

8.2/10
Attack vector Network
Attack complexity Low
Privileges required None
User interaction None
Confidentiality
Integrity

CVSS vector

CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N

What the vulnerability does

01Description

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Key dates

02Disclosure timeline

January 30, 2025 CVE published
February 12, 2025 Record updated

Related vulnerabilities

04Related CVE