CVE-2025-14097 HIGH

CVE-2025-14097: Remote Code Execution Vulnerability in Radiometer Products

Vendor Radiometer Medical Aps
Product ABL90 FLEX and ABL90 FLEX PLUS Analyzers
Weakness CWE-287 · Improper authentication
Published December 17, 2025
Last update December 17, 2025

CVSS base score

7.2/10
Attack vector Network
Attack complexity Low
Privileges required High
User interaction None
Confidentiality High
Integrity High

CVSS vector

CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

What the vulnerability does

01Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Key dates

02Disclosure timeline

December 17, 2025 CVE published
December 17, 2025 Record updated